An analog integrated circuit beamformer for high-frequency medical ultrasound imaging
dc.authorid | 0000-0002-2239-9927 | |
dc.authorid | 0000-0001-9198-8525 | |
dc.contributor.author | Gürün, Gökçe | en_US |
dc.contributor.author | Zahorian, Jaime S. | en_US |
dc.contributor.author | Şişman, Alper | en_US |
dc.contributor.author | Karaman, Mustafa | en_US |
dc.contributor.author | Hasler, Paul E. | en_US |
dc.contributor.author | Değertekin, Fahrettin Levent | en_US |
dc.date.accessioned | 2015-01-15T23:02:03Z | |
dc.date.available | 2015-01-15T23:02:03Z | |
dc.date.issued | 2012-10 | |
dc.department | Işık Üniversitesi, Mühendislik Fakültesi, Elektrik-Elektronik Mühendisliği Bölümü | en_US |
dc.department | Işık University, Faculty of Engineering, Department of Electrical-Electronics Engineering | en_US |
dc.description.abstract | We designed and fabricated a dynamic receive beamformer integrated circuit (IC) in 0.35-mu m CMOS technology. This beamformer IC is suitable for integration with an annular array transducer for high-frequency (30-50 MHz) intravascular ultrasound (IVUS) imaging. The beamformer IC consists of receive preamplifiers, an analog dynamic delay-and-sum beamformer, and buffers for 8 receive channels. To form an analog dynamic delay line we designed an analog delay cell based on the current-mode first-order all-pass filter topology, as the basic building block. To increase the bandwidth of the delay cell, we explored an enhancement technique on the current mirrors. This technique improved the overall bandwidth of the delay line by a factor of 6. Each delay cell consumes 2.1-mW of power and is capable of generating a tunable time delay between 1.75 ns to 2.5 ns. We successfully integrated the fabricated beamformer IC with an 8-element annular array. Experimental test results demonstrated the desired buffering, preamplification and delaying capabilities of the beamformer. | en_US |
dc.description.sponsorship | This work was supported by Award R01EB010070 from the National Institute of Biomedical Imaging and Bioengineering, and Award R01HL082811 from the National Heart, Lung and Blood Institute. This paper was recommended by Associate Editor T. Constandinou. | en_US |
dc.description.version | Publisher's Version | en_US |
dc.identifier.citation | Gurun, G., Zahorian, J. S., Sisman, A., Karaman, M., Hasler, P. E., & Degertekin, F. L. (2012). An analog integrated circuit beamformer for high-frequency medical ultrasound imaging. IEEE Transactions on Biomedical Circuits and Systems, 6(5), 454-467. doi:10.1109/TBCAS.2012.2219532 | en_US |
dc.identifier.doi | 10.1109/TBCAS.2012.2219532 | |
dc.identifier.endpage | 467 | |
dc.identifier.issn | 1932-4545 | |
dc.identifier.issn | 1940-9990 | |
dc.identifier.issue | 5 | |
dc.identifier.pmid | 23853232 | |
dc.identifier.scopus | 2-s2.0-84870563379 | |
dc.identifier.scopusquality | Q1 | |
dc.identifier.startpage | 454 | |
dc.identifier.uri | https://hdl.handle.net/11729/436 | |
dc.identifier.uri | http://dx.doi.org/10.1109/TBCAS.2012.2219532 | |
dc.identifier.volume | 6 | |
dc.identifier.wos | WOS:000313906500007 | |
dc.identifier.wosquality | Q1 | |
dc.indekslendigikaynak | Web of Science | en_US |
dc.indekslendigikaynak | Scopus | en_US |
dc.indekslendigikaynak | PubMed | en_US |
dc.indekslendigikaynak | Science Citation Index Expanded (SCI-EXPANDED) | en_US |
dc.institutionauthor | Karaman, Mustafa | en_US |
dc.language.iso | en | en_US |
dc.peerreviewed | Yes | en_US |
dc.publicationstatus | Published | en_US |
dc.publisher | IEEE-INST Electrical Electronics Engineers Inc | en_US |
dc.relation.ispartof | IEEE Transactions on Biomedical Circuits and Systems | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Engineering | en_US |
dc.subject | All-pass filter | en_US |
dc.subject | Analog delay | en_US |
dc.subject | Beam forming | en_US |
dc.subject | Biquad current mirror | en_US |
dc.subject | Delay line | en_US |
dc.subject | Ultrasound imaging | en_US |
dc.subject | Prml read/write channel | en_US |
dc.subject | Intravascular ultrasound | en_US |
dc.subject | Matrix transducer | en_US |
dc.subject | Annular arrays | en_US |
dc.subject | Cmut arrays | en_US |
dc.subject | Cmos | en_US |
dc.subject | Systems | en_US |
dc.subject | Design | en_US |
dc.subject | Chip | en_US |
dc.title | An analog integrated circuit beamformer for high-frequency medical ultrasound imaging | en_US |
dc.type | Article | en_US |
dspace.entity.type | Publication |
Dosyalar
Orijinal paket
1 - 1 / 1
Küçük Resim Yok
- İsim:
- 436.pdf
- Boyut:
- 2.65 MB
- Biçim:
- Adobe Portable Document Format
- Açıklama:
- Publisher's Version